ALARMS
PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft 119
All information contained herein is subject to the rights and restrictions on the title page.
Table 12. Ambulatory technical alarms
Bedside Monitored Patients
Table 13. Bedside device non-arrhythmia medical alarms
Alarm Label Alarm Triggering Condition
NURSE nurse nurse call button on the transceiver pressed
?? SOURCE source data received does not match configured device type
LOW BATT low transceiver battery battery level on the patient transceiver is below limit
MUSCLE muscle muscle artifact detected
NO SIGNAL no signal no signal received from transceiver
CHKSIGNAL check signal intermittent no signal detected
WRONG ID wrong ID Incompatible transceiver ID
EDIT edit number of history events has reached configured
threshold
LEAD OFF lead off ECG electrode is disconnected or loose
CHECK LEAD check leads One or more of the ECG leads has a poor connection
and/or is causing significant baseline wander
NO ARR Arrhythmia unable to
analyze
No good leads are available for analysis
ATT PRESENT attendant present called when pressed on transceiver
PA SILENCE PAS called when sent from transceiver
Alarm Label Alarm Triggering Condition
HIGH HR
high heart rate heart rate greater than high HR limit
LOW HR
low heart rate heart rate below the low HR limit
HI SpO
2
high SpO
2
high limit exceeded for pulse oximetry
LO SpO
2
low SpO
2
low limit exceeded for pulse oximetry
HI P
1
SYS
high systolic P
1
high limit exceeded for systolic P
1
invasive pressure
LO P
1
SYS
low systolic P
1
low limit exceeded for systolic P
1
invasive pressure
HI P
1
MN
high mean P
1
high limit exceeded for mean P
1
invasive pressure
LO P
1
MN
low mean P
1
low limit exceeded for mean P
1
invasive pressure
HI P
1
DIA
high diastolic P
1
high limit exceeded for diastolic P
1
invasive pressure
LO P
1
DIA
low diastolic P
1
low limit exceeded for diastolic P
1
invasive pressure
HI P
2
SYS
high systolic P
2
high limit exceeded for systolic P
2
invasive pressure
LO P
2
SYS
low systolic P
2
low limit exceeded for systolic P
2
invasive pressure
HI P
2
MN
high mean P
2
high limit exceeded for mean P
2
invasive pressure
LO P
2
MN
low mean P
2
low limit exceeded for mean P
2
invasive pressure
HI P
2
DIA
high diastolic P
2
high limit exceeded for diastolic P
2
invasive pressure
LO P
2
DIA
low diastolic P
2
low limit exceeded for diastolic P
2
invasive pressure