User Manual
ALARMS
112 PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
Re-learning After Alarm State Transitions
Relearning (see “Learn and Relearn Considerations” on page 70) takes place automat-
ically on resumption of monitoring after Smart Alarm and after alarm state transitions
as shown in Table 10.
Table 10. Relearning after alarm state transitions.
CAUTION: The visual display of Learning (LEARNING) in the Patient Tile will not be seen
if V-TACH occurs during a Smart Alarm condition.
From To Re-learn Initiated
Smart Alarm Alarms On Yes
Smart Alarm Alarms Off Yes
Smart Alarm 3 Minute Alarm Off Yes
Smart Alarm Smart Alarm No
Alarms On Alarms On No
Alarms On Alarms Off No
Alarms On 3 Minute Alarm Off Yes
Alarms On Smart Alarm No
Alarms Off Alarms On Yes
Alarms Off Alarms Off No
Alarms Off 3 Minute Alarm Off Yes
Alarms Off Smart Alarm Yes
3 Minute Alarm Off Alarms On Yes
3 Minute Alarm Off Alarms Off Yes
3 Minute Alarm Off 3 Minute Alarm Off N/A
3 Minute Alarm Off Smart Alarm Yes