User Manual
ALARMS
108 PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
When you turn off all alarms, the alarm text in the patient block reads ALARM OFF in
yellow until you click on it.
Medical, Level 2
Less urgent than a medical, level 1 alarm (flashing red, single fast tone).
1. Press the Silence button on the Main screen once.
2. Press the View button and select the alarming patient to observe the patient’s
real-time rhythm in applicable leads.
3. Notify the appropriate staff according to hospital policy and procedure.
4. If the alarm is configured for paging, analyze the rhythm and select Page or
No Page as appropriate.
5. If the alarming rhythm corrects to normal, go to Step 6. If the alarming
rhythm continues, go to Step 7.
6. Select the alarm text in the patient information block to acknowledge and
clear the alarm.
7. Press the Silence button twice and then press the alarm text in the patient
block. This turns off all alarms for 3 minutes.
While the alarms are off, ALARM OFF flashes in yellow. Then, a learn takes place and
all alarms are automatically turned on again after the 3 minute time limit.
If hospital policy allows, when a medical code is in progress and all needed staff are
present in the room, all alarms can be turned off. Do this with the Alarms button in
the Patient Setup screen.
When you turn off an active alarm, the alarm text in the patient block reads “ALARM
OFF” in yellow until you click on it.
Medical, Level 3
Less urgent than a medical, level 2 alarm (flashing red, single slow tone).
1. Press the Silence button on the Main screen once.
2. Press the View button and select the alarming patient to observe the patient’s
real-time rhythm in applicable leads.
3. Notify the appropriate staff according to hospital policy and procedure.
4. If the alarm is configured for paging, analyze the rhythm and select Page or
No Page as appropriate.
5. If the alarming rhythm corrects to normal, go to Step 6. If the alarming
rhythm continues, go to Step 7.
6. Press the Silence button as needed.
7. Press the Silence button twice and then press the alarm text in the patient
block. This turns off all alarms for 3 minutes.
ALARM OFF flashes yellow for 3 minutes, then a learn takes place and the alarms