User Manual
PACEMAKER PROCESSING
100 PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
The following procedure is recommended for determining the length of the optional
blanking interval:
1. Place the patient on the system.
2. After establishing a stable waveform, store the paced rhythm in the view
screen.
3. Inspect the flag and measure the interval from the flag to the physiologic
response (P wave or QRS), as shown in fig. 44 and fig. 45. Residual artifacts
of the pacer may appear between the flag and the start of the QRS. You can
use the electronic calipers to measure the interval.
Fig. 44. Waveform Measurement - Blanking Interval
Fig. 45. Detailed Waveform
40 mm interval
Artifacts
Pacer
Pacer Spike
Interval before the spike
Interval after the spike