PATIENT SETTINGS
PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft 89
All information contained herein is subject to the rights and restrictions on the title page.
Table 9. Waveform labels
Alarms
Information on alarms and instructions on how to set and work with alarms are cov-
ered in detail in the “Alarm Management” chapter starting on page 105.
Enabling Remote Alert
With Remote Alert enabled, a designated PatientNet Viewer automatically displays an
alarming patients channel and sounds the alarm. See “Remote Viewing Stations” on
page 175 for more information.
To enable Remote Alert:
1. Press the Alarm Config button on the Patient Settings screen.
2. Highlight the desired alarm in the alarm configuration screen and toggle the
Assign button to On.
Note: This applies to the PatientNet Viewer and the Interactive-PatientNet Viewer.
Assigning a Pager Number to a Patient
Note: This applies only if your system includes the paging option. See “Paging”
starting on page 238 for further instructions and details on paging.
To assign a pager number or telephone extension to a patient for notification in case of
alarms:
1. Press Pager # on the Patient Settings screen to bring up the on-line numeric
keypad.
2. Enter the pager number or telephone extension (maximum 7 characters) to be
assigned to this patient and press Enter.
3. If the patient belongs to a zone (a group of up to ten different pager numbers),
enter Z and then the zone number. Alarms enabled for paging will be sent to
all pagers in the zone. Zone group paging is set up by the system administra-
tor.
4. Select None if you wish no page to be sent.
Note: Pagers may also be set-up by group, where a number of pagers belong to a
single group.
3-Wire 4-Wire 5-Wire Bedside
Wave 1
N/A I, ML1 I ECG I
Wave 2
I, II, III, ML1,
ML6
II, ML6 II P1
Wave 3
N/A N/A V, V
1
, V
2
, V
3
,
V
4
, V
5
, or V
6
P2
Augment
N/A No, Yes, N/A Yes, No N/A