System Components
WMTS Transceivers
DT-4500
The DT-4500 is an internally powered device and is suitable for continuous operation.
Table 32 DT-4500 Environmental Specifications
Table 33 Device Specifications
Operating Storage
Temperature
+41° F to +104° F
+5° C to +40° C
-4° F to +158° F
-20° C to +70° C
Humidity
15% to 95%
non-condensing
10% to 95%
non-condensing
Pressure/Altitude
1060 hPa to 700 hPa
-1,252’ to 9,840’
1060 hPa to 500 hPa
-1,252’ to 18,280’
Component
Requirements
Environmental Protection
IPX7
Input Configuration
5, electrodes, 1 spare, selectable
Pacemaker Detection
Analog on each lead, independent of sample rate
± 2 mV to ± 700 mV Max (per AAMI EC-13-1992)
Battery
9-Volt alkaline or
9-Volt lithium-Use Kodak Ultralife U9VL or
Energizer L522 Only.
Use of another Lithium battery may present a
risk of fire or explosion. See Warning Section.
Battery Life
9-Volt Alkaline-Minimum, 28.8 hours
Frequency Range Programmable
608-614 MHz
Modulation
Digital encoded FM (FSK)
Dimensions
5.75 in. x 2.50 in. x 1.25 in.
Defibrillator Recover Time
8 Seconds