User Manual
SYSTEM COMPLIANCE
PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft 245
All information contained herein is subject to the rights and restrictions on the title page.
DR-10000 Access Point
UL Classification
The DR-10000 Access Point is classified in accordance with UL 1950, Information
Technology Equipment (including Electrical Business Equipment), and CAN/CSA
C22.2 No. 950-95 (Third Edition).
FCC Compliance in the 174-216 MHz and 608-614 MHz bands
1. The marketing and operation of intentional radiators, under the provisions of
15.242 of the FCC rules, is restricted to biomedical telemetry devices that are
employed on health care facilities premises.
a. A health care facility includes hospitals and other establishments that offer
services, facilities, and beds for use beyond 24 hours in rendering medical
treatment, as well as institutions and organizations that are regularly en-
gaged in providing medical services through clinics, public health facili-
ties, and similar establishments, including governmental entities and
agencies for their own medical activities.
b. This authority to operate does not extend to mobile vehicles, such as am-
bulances, even if those vehicles are associated with a health care facility.
2. The user and the installer of a biomedical telemetry device, which is operating
within the frequency range 174-216 MHz, 470-608 MHz, or 614-668 MHz,
shall ensure that the following minimum separation distances are maintained
between the biomedical telemetry device and the authorized radio services
that are operating on the same frequencies:
c. At least 5.5 km outside of the Grade B field strength contour (64 dBuV/m)
of a TV broadcast station or an associated TV booster station that is oper-
ating within the 470-608 MHz or 614-668 MHz bands.
a. At least 10.3 km outside of the Grade B field strength contour (56 dBuV/
m) of a TV broadcast station or an associated TV booster station that is op-
erating within the 174-216 MHz band.
b. At least 5.1 km outside of the 68 dBuV/m field strength contour of a low
power TV or a TV translator station that is operating within the 174-216
MHz band.
c. At least 3.1 km outside of the 74 dBuV/m field strength contour of a low
power TV or a TV translator station that is operating within the 470-608
MHz or 614-668 MHz bands.
d. Whatever distance is necessary to protect other authorized users within
these bands.