SKIN PREPARATION AND LEAD PLACEMENT
48 PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
Alternative Placement of Electrodes
Surgical Patients
Place the Brown electrode in the area of the V
5
chest lead.
Fig. 15. Surgical Patients Lead Placement
Patients Using Transcutaneous Electrical Nerve Stimulators (TENS)
TENS signals may saturate the signal and can interfere with, or result in gross distor-
tion, of the ECG waveform. This may result in false asystole or low heart rate alarms.
It also affects the displayed ECG waveform. The following interventions are recom-
mended:
Reposition the leads attempting to adjust their position relative to the TENS
unit. Follow the steps detailed for repositioning the leads for patients with
pacemakers (see “Patients with a Pacemaker” below).
Refer to the operator’s manual that is provide by the manufacturer of the
TENS unit.
If TENS signals continue to interfere with ECG monitoring, and if ECG mon-
itoring is critical to patient care, you may need to make the clinical decision to
discontinue TENS pain management.