156 PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
24/72-Hour Summary Report
This report prints summary graphs of all data trended during the last 24/72 hours and
the total number of alarm violations. It leaves space for physician signature and notes.
The data reported depends on the parameters selected for the past 24/72 hours.
The event is listed on the second page if archived, or a report was set.
It is printed at the Central Station even if requested at a PatientNet Viewer.
Zoomed-In Disclosure
Zoomed-in disclosure reports can be printed at either the Central Station or a Patient-
Net Viewer. To print zoomed in waves:
1. Press the mouse button and hold while you highlight (light gray) the segment
of a waveform to zoom in.
2. Press Laser to print all waves in the shaded time period.
Quality Reports
Pressing this button displays the 24 (or 72) Hr Quality popup which allows you to
print a detailed Quality report for either a particular patient or a summary Quality
report for all patients that are on the central station.
Note: This button is disabled when Laser Access is set to No in the system
configuration screen
Patient Button
The 24 hour (or 72-hour) quality report prints the time duration for specific alarm and
history events that are stored in full disclosure over the past 24 (72) hours for the
selected patient. The Date and Time formats on the report match the date and time for-
mats that are configured on the system.
1. The 24-hour (72-hour) quality report prints the following alarm events:
• each lead off alarm duration(s), the average duration leads were off, num-
ber of times the lead off alarm was triggered, and total time the lead off
alarm was active
• alarms off duration(s), the average duration alarms were off, number of
times alarms were set to off, and total time alarms were off
• low battery alarm duration(s), the average duration of the low battery
alarm, number of times the low battery alarm was triggered, and total time
the low battery alarm was active
• no signal alarm duration(s), the average duration of the no signal alarm,
number of times the no signal alarm was triggered, and total time the no
signal alarm was active