User Manual
REMOTE VIEWING STATIONS
178 PatientNet Operator’s Manual, v1.04, 10001001-00X, Draft
All information contained herein is subject to the rights and restrictions on the title page.
How Remote Alert Displays an Alarming Patient
The system assigns the alarming patient to an empty channel on the PatientNet Viewer
(RVS) or Interactive-PatientNet Viewer (IRVS). If there are no empty channels, then
the alarming patient replaces the patient who has been displayed longest on the RVS/
IRVS. In a 16 channel Interactive-PatientNet Viewer in split screen mode the alarming
patient may be placed in a channel not currently visible. In this case the First 8 or
Last 8 button (as appropriate) flashes red.