All information contained herein is subject to the rights and restrictions on the title page.
Programming Transceivers
You can program DT-4500 and DT-7000/7001 transceivers with a programming cable
connected to the I/O board of any Central Station unit.
Required Items
9 V Battery for DT-4500
Power Supply for DT-7000/DT-7001 transceiver
Programming cable for DT-4500 or DT-7000/7001 Transceivers
DT-4500 or DT-7000/DT-7001 transceivers
Attaching Transceiver Programming Cables
For the DT-4500
1. Attach the programming cable to the Port 2 connector on the MSIO board on
the rear of the unit.
2. Remove the battery from the DT-4500 transceiver.
3. Remove and save the port plug, if in place. Connect the other end of the pro-
gramming cable to the connector on the DT-4500 transceiver.
4. Place the 9 V battery into the transceiver, with the power contacts facing
down into the transceiver.
For the DT-7000
1. Attach the programming cable to the Port 3 connector on the MSIO board on
the rear of the unit.
2. Connect the other end of the programming cable to Port 1 on the
DT-7000/DT-7001 transceiver.
3. Connect the DT-7000/DT-7001 power supply to the DT-7000/DT-7001
Power Connector and to the 120 VAC/60 Hz wall power.
Note: The Port numbers listed above are the default ports. If you change the pro-
gramming port parameter in the software, then you must also connect the
programming cable to the appropriate MSIO port based on the selection.