USE AND MAINTENANCE
DT-7000 Operator’s Manual, Revision A 17
All information contained herein is subject to the rights and restrictions on the title page.
Cleaning
This section provides cleaning and maintenance instructions for DT-7000 and DT-
7001 transceivers.
Read and follow all precautions when cleaning transceivers.
WARNING: VitalCom makes no claims concerning the sterilization of the DT-4500
Ambulatory, DT-7000, and DT-7001 Instrument Transceivers.
CAUTION: Do not gas sterilize or AUTOCLAVE any part of the monitoring system,
transmitters, or transceivers.
Note: DO NOT use abrasive cleaners.
Cleaning the Chassis
1. Transceivers can be cleaned with a gauze pad or cloth moistened with one of
the following agents:.
• Soap and Water
• Quaternary Ammonium
• Glutaraldehyde 2%
• Dilute Chlorine Bleach (Sodium hypochlorite), 10% solution, freshly
made in past 24 hours
• Isopropyl Alcohol 70%
• Ethyl Alcohol
2. Use a cloth moistened with distilled water to rinse away the cleaning solution.
3. Dry thoroughly with a lint-free cloth.