User Manual
38
You can delete this control test from memory, if desired. For more
information, refer to the Dario App User Guide at
www.mydario.com.
6. Compare the results on the meter with the values displayed on the test
strip cartridge. If the results are:
• Within the acceptable target values, then your meter is ready for
use.
• Outside of the acceptable target values, perform the test again with
a new test strip. If the problem recurs, contact customer support
toll-free in the U.S. at 1-800-895-5921, 9:00 AM-5:00 PM EST
Monday through Friday. Internationally visit www.mydario.com.
7. When the test is done, remove the test strip and follow your health care
provider’s instructions for disposal.
8. Disconnect the meter from your smart mobile device.
9. Wipe the tip of the control solution container clean and tightly seal the
container.
The Dario Control Solution target values printed on the test strip
cartridge label are intended for the Dario Test Strip Cartridge only.
When measured by systems other than the Dario LC Blood
Glucose Monitoring System, the glucose concentrations in these
solutions may differ.
Applying too much or too little control solution to the test strip
may cause out-of-range results.
Do not use your Dario LC Blood Glucose Monitoring System to
test your blood if your control solution result falls outside of the
acceptable target values.
The Dario LC Blood Glucose Monitoring System includes a
function for specific control solution tests. Control solution tested
values must be tested under the "Control solution test" function:
!
!
!
!