© National Instruments Corporation 4-17 NI 6034E/6035E/6036E User Manual
Figure 4-7 shows how to connect a grounded signal source to a channel
configured for NRSE mode on the NI 6034E/6035E/6036E device.
Figure 4-7. Single-Ended Input Connections for Ground-Referenced Signals
Common-Mode Signal Rejection Considerations
Figures 4-4 and 4-7 show connections for signal sources that are
already referenced to some ground point with respect to the
NI 6034E/6035E/6036E device. In these cases, the PGIA can reject any
voltage caused by ground potential differences between the signal source
and the device. In addition, with differential input connections, the PGIA
can reject common-mode noise pickup in the leads connecting the signal
sources to the device. The PGIA can reject common-mode signals as long
as V+
in
and V
in
(input signals) are both within ±11 V of AIGND.
Selected Channel in NRSE Configuration
PGIA
Programmable
Gain
Instrumentation
Amplifier
+
+
+
Input Multiplexers
AISENSE
AIGND
V
cm
+
V
s
ACH+
ACH
V
m
Common-
Mode
Noise and
Ground
Potential
Ground-
Referenced
Signal
Source
Measured
Voltage
I/O Connector