Technical data

5.1.5 Audit log report
The Daniel 3810 Series Liquid Ultrasonic Flow Meter stores an audit log record whenever
any parameter affecting the liquid flow measurement is modified. The audit log record
indicates which data point changed, the date and time of the change, and both the “As-
found” and “As-left” values.
The meter can store up to 3000 audit records. The user can select whether old, unread
records can be overwritten by new records when the log becomes full via the data point
DoOverwriteUnreadAuditLog. This point can be modified using the Daniel MeterLink Tools -
Edit/Compare Configuration screen. The default is to overwrite old, unread records. Refer to
Section 5.2.5 for information on reading records and marking records as read. The data
point IsAuditLogFull indicates whether or not the audit log is full and cannot overwrite old,
unread records.
The data points monitored for the audit log are as shown in (Table 5-2). The points are
grouped and, within each group, are listed alphabetically. The groupings are as follows:
Calibration
Chord Proportions
Co-located meter
Communications
Data Logging
Expansion Correction
Flow analysis
Frequency, Digital, and Analog Signals
General
HART
Indicators
Local display
Meter Information
Pressure and Temperature
Reynolds Number Calculation
Signal Processing
Tracking
Audit log data
Group monitored data points
Audit log calibration group monitored data pointsTable 5-2:
Calibration data group data points
AvgDlyA
AvgDlyB
Archive logs and reports
Operations 69