Technical data

HART outputsFigure 4-13:
1. Click the HART outputs checkbox on the Field Setup Wizard Startup page.
2.
Use the pull-down menu to select the Process variables,
Primary variable - Uncorrected flow rate
Secondary variable - Uncorrected flow rate
Third variable - Uncorrected flow rate, pressure or temperature
Fourth variable - Uncorrected flow rate, pressure or temperature
3. Input the device identification parameters,
Tag - the tag name for the HART device which may be used by host systems to
uniquely identify the meter. The tag may be up to 8 characters in length.
Long tag - the specified long tag matches that of the meter. The tag may be 32
bytes in length.
Date - a 3 byte value where the most significant byte is the day of the month
(1-31), the second byte is the month of the year (1-12), and the third byte is the
year-1900.
Message - a string value that can be no more than 32 characters in length.
Descriptor - a string value that can be no more than 16 characters in length.
Final assembly number - a numeric value that must be between 0 and 1677215.
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
52 Daniel 3810 Series Operations Manual