Technical data
• Digital output 1B is based on Digital output 1B content (Frequency 1B Validity and
Flow Direction)
• Digital output 2A is based on Digital output 2A content (Frequency 2A Validity and
Flow Direction)
• Digital output 2B is based on Digital output 2B content (Frequency 2B Validity and
Flow Direction)
Frequency or Digital Outputs (FODO 3) source
• FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
• FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
• Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate
or Corrected Flow Rate)
• Frequency output 1B is based on frequency content and Frequency 1B Phase
• Frequency output 2A is based on frequency content (Actual - Uncorrected Flow
Rate)
• Frequency output 2B is based on frequency content and Frequency 2B Phase
• Digital output 1A is based on Digital output1A content (Frequency 1A Validity and
Flow Direction)
• Digital output 1B is based on Digital output 1B content (Frequency 1B Validity and
Flow Direction)
• Digital output 2A is based on Digital output 2A content (Frequency 2A Validity and
Flow Direction)
• Digital output 2B is based on Digital output 2B content (Frequency 2B Validity and
Flow Direction)
Mode options
• Open Collector (requires external excitation supply voltage and pull-up resistor)
• TTL (internally powered by the meter 0-5 VDC signal)
Channel B Phase options
• Lag forward, Lead reverse (Phase B lags Phase A while reporting forward flow, leads
Phase A while reporting reverse flow)
• Lead forward, Lag reverse (Phase B leads Phase A while reporting forward flow, lags
Phase A while reporting reverse flow)
Phase A and Phase B output (based on flow direction)
• Reverse flow - output only reports flow in the reverse direction. For frequency
outputs, Phase B of the output is 90 degrees out of phase with Phase A.
• Forward flow - output only reports flow in the forward direction. For frequency
outputs, Phase B of the output is 90 degrees out of phase with Phase A.
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
Operations 45