Technical data

Digital output 1B is based on Digital output 1B content (Frequency 1B Validity and
Flow Direction)
Digital output 2A is based on Digital output 2A content (Frequency 2A Validity and
Flow Direction)
Digital output 2B is based on Digital output 2B content (Frequency 2B Validity and
Flow Direction)
Frequency or Digital Outputs (FODO 3) source
FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate
or Corrected Flow Rate)
Frequency output 1B is based on frequency content and Frequency 1B Phase
Frequency output 2A is based on frequency content (Actual - Uncorrected Flow
Rate)
Frequency output 2B is based on frequency content and Frequency 2B Phase
Digital output 1A is based on Digital output1A content (Frequency 1A Validity and
Flow Direction)
Digital output 1B is based on Digital output 1B content (Frequency 1B Validity and
Flow Direction)
Digital output 2A is based on Digital output 2A content (Frequency 2A Validity and
Flow Direction)
Digital output 2B is based on Digital output 2B content (Frequency 2B Validity and
Flow Direction)
Mode options
Open Collector (requires external excitation supply voltage and pull-up resistor)
TTL (internally powered by the meter 0-5 VDC signal)
Channel B Phase options
Lag forward, Lead reverse (Phase B lags Phase A while reporting forward flow, leads
Phase A while reporting reverse flow)
Lead forward, Lag reverse (Phase B leads Phase A while reporting forward flow, lags
Phase A while reporting reverse flow)
Phase A and Phase B output (based on flow direction)
Reverse flow - output only reports flow in the reverse direction. For frequency
outputs, Phase B of the output is 90 degrees out of phase with Phase A.
Forward flow - output only reports flow in the forward direction. For frequency
outputs, Phase B of the output is 90 degrees out of phase with Phase A.
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
Operations 45