Technical data
Analog outputs data points (continued)Table 4-7:
Daniel Meter-
Link Display
Name Data Points, Options and Guidelines
Current Output
Start/StopAO2
Data points affected:
IsAO2EnableTest
Options:
Click StartAO2 to enter the test mode (TRUE) (Note: The Start button will be-
come the Stop button after clicking on it.)
Click StopAO2 to exit the test mode (FALSE) (Note: the Stop button will be-
come the Start button after clicking on it.) Guidelines: The specified test value
takes effect within one batch period of clicking the Start button.
Configure Frequency/Digital output sources
The meter has three user-configurable selections for configuring either a Frequency
output or Digital output (FODO).
• FODO1 (four possible parameter configurations)
• FODO2 (eight possible parameter configurations)
• FODO3 (eight possible parameter configurations
Frequency or Digital Outputs (FODO 1) source
• FO1A, DO1A, FO1B, DO1B
• Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate
or Corrected Flow Rate)
• Frequency output 1B is based on frequency content and Frequency 1 B Phase
• Digital output 1A is based on Digital output1A content (Frequency1A Validity and
Flow Direction)
Frequency or Digital Outputs (FODO 2) source
• FO1A, DO1A, FO1B, DO1B, FO2A, DO2A, FO2B, DO2B
• Frequency output 1A is based on frequency content (Actual - Uncorrected Flow Rate
or Corrected Flow Rate)
• Frequency output 1B is based on frequency content and Frequency 1B Phase
• Frequency output 2A is based on frequency content (Actual - Uncorrected Flow
Rate)
• Frequency output 2B is based on frequency content and Frequency 2B Phase
• Digital output 1A is based on Digital output1A content (Frequency 1A Validity and
Flow Direction)
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
44 Daniel 3810 Series Operations Manual