Technical data

4. Enter the correct value in the Actual edit box. Finally, the Finish screen shows the new
calculated offset and gain values
5.
Click Finish to write the values to the meter. If the meter is not configured to use live
values from the analog input for calculation, a prompt displays asking if the
configuration should be changed to use the live input.
6. Place the WRITE PROT. switch on the CPU Module to ON to write-protect the meter's
configuration.
Analog output data points
Analog outputs data pointsTable 4-7:
Daniel Meter-
Link Display
Name Data Points, Options and Guidelines
Current output
Output setting
Data points affected:
AO1TestModeOutputPercent
Options:
Enter the integer percentage of full scale for the analog output within the
range [0, 100 %] where 0% corresponds to the minimum output (4 mA)
and 100% corresponds to the maximum output (20 mA).
Guidelines:
The specified test value takes effect within one batch period of clicking the
Start button.
Current output
Start(Stop)AO1
Data points affected:
IsAO1EnableTest
Options:
Click StartAO1 to enter the test mode (TRUE) (Note: The Start button will
become the Stop button after clicking on it.)
Click StopAO1 to exit the test mode (FALSE) (Note: The Stop button will
become the Start button after clicking on it.)
Guidelines:
The specified test value takes effect within one batch period of clicking the
Start button.
Current output
Output setting
Data points affected:
AO2TestModeOutputPercent
Options:
Enter the integer percentage of full scale for the analog output within the
range [0, 100 %] where 0% corresponds to the minimum output (4 mA)
and 100% corresponds to the maximum output (20 mA).
Guidelines:
The specified test value takes effect within one batch period of clicking the
Start button.
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
Operations 43