5. Click Save to open the Save As dialog box.
A default name is provided based on the meter name, time, and date the
configuration was collected. By default, all files are saved in the Data folder directory
defined under Program Settings. You may change the name and directory location if
desired. The entire configuration is saved to file.
4.1.9 Calibrate analog inputs and outputs
Using Daniel MeterLink
™
Analog Inputs Wizard
An analog input can be calibrated regardless of the corresponding input type, pressure or
temperature. However, if the corresponding input type is selected as Live, then the input
being calibrated is considered to be invalid and the in-use value depends upon the
FlowPOrTSrcUponAlarm
selection (see above). In this case, the appropriate data point
(FlowPressureWhileCal, FlowTemperatureWhileCal) is set to the in-use value so that it can
be logged in the optional log.
The offset and gain can be reset to the default values (0 and 1, respectively) by clicking the
Reset Defaults button.
Important
Clicking the Reset Defaults button writes the offset and gain immediately and the previous values
cannot be restored.
Click the Edit Scaling button to modify the input scaling.
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters