Technical data
2. Start the calibration by clicking Start Zero. The calibration will start and run for the
period of time set in Zero time.
The average flow velocity is the value used to adjust the FwdA0 and RevA0
coefficients in the meter so they remove the offset flow under no flow conditions.
3.
The Estimated maximum deviation is also calculated to help verify the meter is in no
flow and is used to set pass/fail criteria for the calibration.
An error will be generated if the Estimated maximum deviation is greater than 0.002
ft/sec or 0.0006 m/sec.
4. Once the calibration completes and no errors were encountered, click Write to Meter
to send the changes to the meter.
If errors were reported after the calibration, the calibration cannot be written to the
meter
5. If you selected to generate a report, Daniel MeterLink displays a dialog asking you if
you want to display the report once created.
6. Click Stop Zero, at any time during a calibration, to abort the calibration.
7. Click Discard once a calibration completes with no errors, to cancel the calibration
and setup the dialog to perform a new calibration.
8. Click Close to close the Zero Calibration dialog
4.1.7 Triggered Delta Volumes
The "triggered delta volume" feature provides the ability to measure total liquid flow
volume between two successive external event triggers.
To trigger an event, set the DoUpdtTrigDeltaVols data point to TRUE. This causes the
meter to save the current accumulated flow- and base-condition volume values (forward
and reverse). The meter then calculates the difference between these values and the
corresponding values saved from the previous event trigger. Finally the meter writes the
delta volume values to the appropriate data points (TrigDeltaPosVolFlow and
TrigDeltaNegVolFlow) and sets the DoUpdtTrigDeltaVols data point to FALSE to clear the
trigger and indicate the calculation completion
The triggered delta volume functionality is retained across power cycles as the
accumulated volume values at the last event trigger are saved in non-volatile memory.
The delta volume data points are stored internally in non-volatile memory as double-
precision floating point numbers. The delta volume data points can be read via Modbus as
either 32-bit floating point values or as integer values (using the [overflow, lower] LONG
register pair in a manner similar to reading the accumulated volumes).
Suggested User Access Logic
The following pseudo-code demonstrates the logic to access the triggered delta volume
functionality:
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
Operations 39