Technical data
4.1.4 Calibrate the meter factors
Calibrate Meter factor parameters
1.
Use the Calibration¦Meter Factors menu and select Flow Calibration Method¦Meter Factor from
the pull-down menu.
2. Enter the appropriate forward and reverse meter factors and click Write to send the
factors to the meter.
3. For High Viscosity applications, use the Calibration¦Meter Factors menu and select Flow
Calibration Method¦Meter Factor from the pull-down menu
• Select High Viscosity Method Enabled from the pull-down menu. See Section 4.1.3
for additional configuration parameters.
4. Click Close to exit the dialog box. If you made changes, click the Write button to
apply the changes.
5. After the meter is configured, set the WRITE PROT. switch on the CPU Module to
the ON position to write-protect the meter's configuration.
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
34 Daniel 3810 Series Operations Manual