Technical data

Select the desired Customer Calibration by configuring the data point listed in the
Table 4-2; from the Daniel MeterLink Calibration - Meter Factors page.
Customer CalibrationTable 4-3:
Daniel MeterLink Display Name Data Points, Options and Guidelines
Flow calibration method Data points affected: CalMethod
Options:
None (0)
Meter Factor (3)
High Viscosity Method (Disabled - default)
(Enabled)
Guidelines: This value should only be modified
under direction of Daniel personnel.
Should your high viscosity meter need to be re-calibrated, contact Daniel Customer
Service. Refer to the Field Calibration of High Viscosity, Low Reynolds Number Ultrasonic Meter
Assemblies Procedure (Daniel P/N P-02871).
4.1.3 High viscosity calibration method
If High Viscosity Method is set to “Enabled”, then the linear flow velocity is a result of
applying the zero calibration high viscosity flow offset (FwdA0HighViscosity or
RevA0HighViscosity) and the high viscosity piecewise linearization meter factor
(LinearMtrFctrHighViscosity) to the average weighted flow velocity (AvgWtdFlowVel).
Configure High Viscosity MethodFigure 4-1:
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
32 Daniel 3810 Series Operations Manual