Technical data

High viscosity calibration coefficients (continued)Table 4-2:
Daniel MeterLink Display Name Data Points, Options and Guidelines
Forward A1 High Viscosity Data points affected:
FwdA1High Viscosity
Options:
Enter a value (dimensionless) within the range
[0.95,1.05].
Guidelines:
This value should only be modified under direction
of Daniel personnel.
Forward A2 High Viscosity Data points affected:
FwdA2High Viscosity
Options:
Enter a value (s/m or s/ft) within the range [0.1,0.1
s/m].
Guidelines:
This value should only be modified under direction
of Daniel personnel.
Forward A3 High Viscosity Data points affected:
FwdA3High Viscosity
Options:
Enter a value (s
2
/m
2
or s
2
/ft
2
) within the range
[0.1,0.1 s
2
/m
2
].
Guidelines:
This value should only be modified under direction
of Daniel personnel.
Reverse A0 High Viscosity Data points affected: Data points affected:
RevA0
Options:
Enter a value (m/s or ft/s) within the range [1,1 m/
s].
Guidelines:
This value should only be modified under direction
of Daniel personnel.
Reverse A1 High Viscosity Data points affected: Data points affected:
RevA1High Viscosity
Options:
Enter a value (dimensionless) within the range
[0.95,1.05].
Guidelines:
This value should only be modified under direction
of Daniel personnel.
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
30 Daniel 3810 Series Operations Manual