Technical data
High viscosity calibration coefficients (continued)Table 4-2:
Daniel MeterLink Display Name Data Points, Options and Guidelines
Forward A1 High Viscosity Data points affected:
• FwdA1High Viscosity
Options:
• Enter a value (dimensionless) within the range
[0.95,1.05].
Guidelines:
• This value should only be modified under direction
of Daniel personnel.
Forward A2 High Viscosity Data points affected:
• FwdA2High Viscosity
Options:
• Enter a value (s/m or s/ft) within the range [‑0.1,0.1
s/m].
Guidelines:
• This value should only be modified under direction
of Daniel personnel.
Forward A3 High Viscosity Data points affected:
• FwdA3High Viscosity
Options:
• Enter a value (s
2
/m
2
or s
2
/ft
2
) within the range
[‑0.1,0.1 s
2
/m
2
].
Guidelines:
• This value should only be modified under direction
of Daniel personnel.
Reverse A0 High Viscosity Data points affected: Data points affected:
• RevA0
Options:
• Enter a value (m/s or ft/s) within the range [‑1,1 m/
s].
Guidelines:
• This value should only be modified under direction
of Daniel personnel.
Reverse A1 High Viscosity Data points affected: Data points affected:
• RevA1High Viscosity
Options:
• Enter a value (dimensionless) within the range
[0.95,1.05].
Guidelines:
• This value should only be modified under direction
of Daniel personnel.
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
30 Daniel 3810 Series Operations Manual