Technical data
Factory calibration coefficients (continued)Table 4-1:
Daniel MeterLink Display Name Data Points, Options and Guidelines
Reverse A1 Data points affected: Data points affected:
• RevA1
Options:
• Enter a value (dimensionless) within the range
[0.95,1.05].
Guidelines:
• This value should only be modified under direction
of Daniel personnel.
Reverse A2 Data points affected:
• RevA2
Options:
• Enter a value (s/m or s/ft) within the range [‑0.1,0.1
s/m].
Guidelines:
• This value should only be modified under direction
of Daniel personnel.
Reverse A3 Data points affected:
• RevA3
Options:
• Enter a value (s
2
/m
2
or s
2
/ft
2
) within the range
[‑0.1,0.1 s
2
/m
2
].
Guidelines:
• This value should only be modified under direction
of Daniel personnel.
High viscosity calibration method
If the high viscosity calibration method selector (HighViscosityMethod) is “Enabled”, then
the result of applying the zero calibration high viscosity flow offset is applied to the
average weighted flow velocity (shown in the following table).
High viscosity calibration coefficientsTable 4-2:
Daniel MeterLink Display Name Data Points, Options and Guidelines
Forward A0 High Viscosity Data points affected:
• FwdA0HighViscosity
Options:
• Enter a value (m/s or ft/s) within the range [‑1,1 m/
s].
Guidelines:
• This value should only be modified under direction
of Daniel personnel.
Daniel Liquid Ultrasonic Flow Meters calibration and configuration parameters
Operations 29