Technical data
Waveform chart keyboard commands (continued)Table 6-2:
Function Keystroke Description
Toggle Nearest Point Tip Ctrl+F9 Turns on tool tip showing the coor-
dinates of the nearest data point to
the mouse pointer.
Print Ctrl+P Prints the displayed chart.
Copy to clipboard Ctrl+C Copies the displayed chart to the
Windows
®
clipboard as table data.
Paste from clipboard Ctrl+V Paste data from the Windows
®
clipboard to the chart utility. The
data must be in the appropriate
format to correctly paste as a new
series to the chart. Copy data from
the chart to a text file to see the ap-
propriate format.
Waveform Zoom Ctrl+I Turns on/off the zoom feature
while in Waveform Read or Stream
to file mode.
6.1.3 Outputs test
Tools menu
The Outputs Test dialog box allows you to monitor the live values of all the frequency,
current and digital outputs. Additionally the outputs can be set into a Test mode to force
the outputs to a specific user defined value. This dialog box is only available while
connected to a meter.
Refer to Outputs test mode in this manual for additional information about the Outputs
Test Mode.
6.1.4 Transducer swap-out
The transducer swap-out utility allows you to update parameters such as path lengths,
delay times, and delta times for chord. This is necessary anytime transducers, mounts,
holders, or stalks have to be replaced for a chord. Refer to the Liquid Ultrasonic Flow Meter
maintenance manuals (P/N 3-9000-764 and P/N 3-9000-767) to replace the transducers in
the field.
Component Selection
The Transducer Swap-out Wizard Current Chord Parameters page displays the first chord
selected. Under Select which components have changed for chord (A, B, C, or D) and select which
components were changed for the chord.
Tools commands utility
106 Daniel 3810 Series Operations Manual