3.7.3 Selecting the Calibration Sample Size
Setting the Number of Frames to Sample
Purpose: Sets the number of frames to sample when performing pixel
coefficient calculations. Higher values cause calibration to take
longer but provide the most accurate results.
Syntax:
css i
Syntax Elements:
i
Number of lines to sample. Allowable values are 32, 64, 128
(factory setting), 256, 512, or 1024.
Notes:
To return the current setting, use the gcp command.
Example:
css 1024
3.7.4 Performing FPN Calibration
Calibrating All Camera Pixels
Purpose: Performs FPN calibration and eliminates FPN noise by subtracting
away individual pixel dark current.
Syntax:
ccf
Notes:
Before performing this command, stop all light from entering the
camera. (Tip: cover lens with a lens cap.)
The goal is to subtract all non-uniformities and offsets in order
to obtain a 0DN output in the dark. Analog offset should
therefore be set to 0 since it gets subtracted out during CCF.
Set the digital gain (ssg) to X1 since during calibration it gets
forced to X1
Perform FPN correction before PRNU correction.
The ccf command is not available when the camera is using the
factory calibrated coefficients (csn 0-7). You must select the
user coefficient set (csn 8-15) before you can perform FPN
calibration. An error message is returned if you attempt to
perform FPN calibration when using csn 0-7.
Example:
ccf