Calibrating Individual Pixels
S
Syntax
ts:
x
Purpose: ets an individual pixelโs PRNU coefficient.
spc x y i
Syntax Elemen
The pixel column number from 1 to 2352.
y
1728. The pixel row number from 1 to
i
where Coefficient value in a range from 0 to 12287
PRNU coeff = 1 + (
4096
i
)
Notes:
the camera is using
csn 0). You must select the
ser coefficient set (
csn 1) before you can perform PRNU
g
csn 0. To return the
csn get csn.
๎ The spc command is not available when
the factory calibrated coefficients (
u
calibration. An error message is returned if you attempt to
perform PRNU calibration when usin
current
number, send the command
Example:
spc 10 50 500
t PRNU Coefficients
Saves the current PRNU coefficients to non-v
Sy
wpc
coefficient set (csn 1) before you c
perform PRNU calibration when using csn 0
current
csn number, send the command ge
wpc
t FPN Coefficients
Pu
u m
3.6.6 Saving, Loading and Resetting Coefficients
Saving the Curren
Purpose: olatile memory.
ntax:
Notes:
๎ The wpc command is not available when the camera is using the
factory calibrated coefficients (csn 0). You must select the user
an perform PRNU
calibration. An error message is returned if you attempt to
. To return the
t csn.
Example:
Saving the Curren
rpose: Saves the current FPN coefficients to non-volatile memory.
Syntax:
wfc
Notes:
๎ The wfc command is not available when the camera is using the
factory calibrated coefficients (csn 0). Yo ust select the user
coefficient set (csn 1) before you can save FPN coefficients. An
error message is returned if you attempt to save FPN coefficients
when using csn 0. To return the current
csn number, send the
command
get csn.
Example:
wfc
DALSA 03-32-10121-09