User`s manual

54 Piranha HS 4x User’s Manual
03-32-10040-03 PRELIMINARY DALSA Corp.
B1 All Available Commands
Table 14: All Available Commands
Command
Function
Syntax Parameters Command and Parameter
Description
calibrate analog
gain
cag t i
Calibrates the analog offset.
t = Tap value. 0 for all taps or 1-4 for
individual tap selection.
i= Line average in a range dependent on
the current camera data mode.
64 to 251
DN for 8 bit mode or 256 to 1007 DN for
10 bit mode.
calibrate analog
offset
cao t i
Calibrates the analog offset.
t = Tap value. 0 for all taps or 1-4 for
individual tap selection.
i= Target value from 1 to 100 DN for 8 bit
mode and 4 to 400 DN for 10 bit mode.
correction
calibrate FPN
ccf [i]
Start FPN coefficient calibration. Use css
to set sample size. Values range from
1 to
100 DN for 8-bit mode and 4 to 400 DN
for 10-bit.
correction
calibrate PRNU
ccp [i]
Start PRNU coefficient calibration. Use css
to set sample size. Values range from 64 to
251 DN for 8-bit mode and 256 to 1007
DN for 10-bit.
correction set
sample
css i
Sets the number of line samples averaged
when using the
gl and gla commands.
i= 16, 32, 64 (factory setting)
display pixel
coefficients
dpc [i i]
Display the pixel coefficients in the order
FPN, PRNU, FPN, PRNU, … Optional
pixel start and end values in a range from 1
to sensor pixel count.
end of line
sequence
els i
Sets whether to use an end of line sequence
where
i is:
0 = End of line sequence disabled
1 = End of line sequence enabled (factory
setting)
enable pixel
coefficients
epc i i
Sets whether pixel coefficients are enabled
or disabled.
The first parameter sets the FPN
coefficients where i is:
0 = FPN coefficients disabled
1 = FPN coefficients enabled
The second parameter sets the PRNU
coefficients where i is:
0 = PRNU coefficients disabled
1 = PRNU coefficients enabled
Parameters:
t = tap id
i = integer value
f = real number
x1 = pixel start
number
x2 = pixel end
number
[ ] = optional