User`s manual
Piranha HS 4x User’s Manual 37
DALSA Corp. PRELIMINARY 03-32-10040-03
Setting Thresholds
To set a lower threshold value that is checked for and reported in the end-of-line statistic, use
the command:
Syntax:
slt i
Syntax Elements:
i
Lower threshold value. Available values are 0 to 255 for 8 bit
more or 0 to 1023 for 10 bit mode.
Example:
slt 550
To set an upper threshold value that is checked for and reported in the end-of-line statistic, use
the command:
Syntax:
sut i
Syntax Elements:
i
Upper threshold value. Available values are 0 to 255 for 8 bit
mode or 0 to 1023 for 10 bit mode.
Example:
sut 1000
3.13 Monitoring Tasks
The camera enters a warning state when any of the camera's continuously running
monitoring tasks detect a failure. Use the
wed command to display the status of all the
defined monitory tasks (if no parameter is passed) and/or to enable/disable specific
monitoring tasks. Table 9 below lists the monitoring tasks.
Table 9: Piranha HS 4x Monitoring Tasks
Parameters Monitoring Task Description
0 0
Disables all monitoring tasks.
0 1
All
Enables all monitoring tasks.
1 0
Monitors all camera voltages.
1 1
Voltage
Disables monitoring of camera voltages.
2 0
Monitors camera temperature.
2 1
Temperature
Disables monitoring of camera
temperature.
3 0
Monitors the external sync signal.
3 1
External Sync
Disables monitoring of the external sync
signal.
4 0
Monitors current setting of analog gain.
4 1
Gain out of spec
Disables monitoring of analog gain
setting.
Note: By default, all
monitoring tasks are
disabled.