User`s manual

Piranha HN RoHS User Manual
Teledyne DALSA 03-032-20135-00
43
Performing FPN Correction
Syntax: Performs FPN correction and eliminates FPN noise by subtracting away individual
pixel dark current. For a complete description on how to use this command, see the Flat
Field Correction Overview.
Syntax:
ccf
Notes:
Before performing this command, stop all light from entering the camera. (Tip:
cover lens with a lens cap.)
Perform FPN correction before PRNU correction.
The ccf command is not available when the CCD direction is externally
controlled (scd 2) (see
Direction Control). Direction control must be stable while
the camera is calculating coefficients.
Available in TDI Mode only.
Save coefficients before changing directions, changing operating mode, or
powering off.
Related Commands:
cpa
Example:
ccf
Setting a Pixel’s FPN Coefficient
Purpose: Sets an individual pixel’s FPN coefficient.
Syntax
sfc x i
Syntax Elements:
x
The pixel number from 1 to sensor pixel count.
i
Coefficient value in a range from 0-511 (12-bit LSB).
Notes:
Available in TDI Mode only.
Example:
sfc 10 50
Returning FPN Coefficients
Purpose: Returns a pixel’s FPN coefficient value in DN (12-bit LSB)
Syntax:
gfc i
Syntax Elements:
i
The pixel number to read in a range from 1 to sensor pixel count.
Notes:
Available in TDI Mode only.
Example:
gfc 10
Performing PRNU to a user entered value
Purpose: Performs PRNU calibration to user entered value and eliminates the difference in
responsivity between the most and least sensitive pixel creating a uniform response to light.
Using this command, you must provide a calibration target.
Executing these algorithms causes the ssb, sag and sab commands to be set to 0 (no
background addition or subtraction, and unity system gain) and the ssg command to 4096
(unity digital gain).
Syntax:
cpa m i
Syntax Elements:
m
PRNU calibration algorithm to use: