User`s manual
Linea 2K, 4K, 8K and 16K Camera User's Manual  PRELIMINARY    85 
Teledyne DALSA    03-032-20206-01 
Cleaning the Sensor Window 
Recommended Equipment 
  Glass cleaning station with microscope within clean room. 
  3M ionized air gun 980 
(http://solutions.3mcanada.ca/wps/portal/3M/en_CA/WW2/Country/) 
  Ionized air flood system, foot operated. 
  Swab (HUBY-340CA-003) 
(http://www.cleancross.net/modules/xfsection/article.php?articleid=24) 
  Single drop bottle (FD-2-ESD) 
  E2 (Eclipse optic cleaning system (www.photosol.com) 
Procedure 
  Use localized ionized air flow on to the glass during sensor cleaning. 
  Blow off mobile contamination using an ionized air gun. 
  Place the sensor under the microscope at a magnification of 5x to determine the 
location of any remaining contamination. 
  Clean the contamination on the sensor using one drop of E2 on a swab. 
  Wipe the swab from left to right (or right to left but only in one direction). Do 
this in an overlapping pattern, turning the swab after the first wipe and with 
each subsequent wipe. Avoid swiping back and forth with the same swab in 
order to ensure that particles are removed and not simply transferred to a new 
location on the sensor window. This procedure requires you to use multiple 
swabs. 
  Discard the swab after both sides of the swab have been used once. 
  Repeat until there is no visible contamination present. 