Specifications

5-16 Performance Tests
DS360 Ultra Low Distortion Function Generator
b) Press
[] (function down) to select 2-Tone.
[FREQ] [1] [kHz]
[AMPL] [1] [V
RMS
]
[SHIFT] [T1/T2]
[FREQ] [2] [0] [kHz]
[AMPL] [.] [1] [V
RMS
]
c) Span the spectrum analyzer down to measure the IMD products (2T1+T2...) for the first 2-
Tone, at the frequencies listed in the table. Record the results on the test record.
d) Press
[SHIFT] [T1/T2]
[FREQ] [1] [0] [kHz]
[AMPL] [1] [V
RMS
]
[SHIFT] [T1/T2]
[FREQ] [1] [2] [kHz]
[AMPL] [1] [V
RMS
]
d) Span the spectrum analyzer down to measure the IMD products (2T1+T2...) for the second 2-
Tone, at the frequencies listed in the table. Record the results on the test record.
5) The square wave rise time is measured using the scope.
a) Disconnect the DS360 from the spectrum analyzer. Connect the + BNC of the DS360 to CH A
of the scope. Set the scope CH A to 2µs/div, 5 V/div, with triggering on CH A.
b) Press
[] (function up) until square wave is selected.
[FREQ] [1] [0] [0] [kHz]
[AMPL] [1] [0] [V
RMS
]
c) Measure the 10% - 90% rise time of the squarewave. Repeat for the fall time. Record the
results on the test record.
d) Connect the - BNC of the DS360 to the scope CH A and repeat step 5c) for the - channel.
6) This completes the waveform tests.