Specifications

5-20 Performance Tests
DS360 Ultra Low Distortion Function Generator
d) Measure the output voltage on the DVM. Record the results.
4) Measure the residule offset (0 V
DC
) for square waves, white noise and pink noise.
a) Press
[] (function down) to select square waves.
[AMPL] [1] [V
RMS
]
b) Measure the output voltage on the DVM. Record the results.
c) Disconnect the sync signal from the DS360 to the DVM. Set the integration time to the longest
possible, to ensure accurate DC measurements of the noise signals.
d) Press
[] (function down) to select white noise.
[AMPL] [1] [V
RMS
]
e) Measure the output voltage on the DVM. Record the results.
f) Press
[] (function down) to select pink noise.
[AMPL] [1] [V
RMS
]
g) Measure the output voltage on the DVM. Record the results.
5) This completes the offset tests.