User's Manual

The PIN code can be set in this menu.
The settable value range:
- 0000 ~ 9999.
* To change digits of PIN code, press the power button for 2 seconds.
* To raise the number of PIN code, press the power button shortly.
* To reduce the number of PIN code, press the menu button shortly.
* To save the setting and return to the advanced menu list, press and hold the power button and
the menu button together for 2 seconds.
* To return to advanced menu list without saving, press the menu button for 2 seconds.
8. Specification
Indication of B.A.C
0.000 ~ 0.500 %BAC
0.00 ~ 5.00
0.00 ~ 2.50 mg/L
Accuracy
±0.005 %BAC (at 0.100 %BAC, at 25 ℃)
±0.05 ‰ (at 1.00 , at 25 ℃)
±0.025 mg/L (at 0.50 mg/L, at 25 ℃)
Warm up time
Within 10 seconds at 0.050 %BAC
* Warm up time may vary depending on the BAC measured
Sensor type
Premium Fuel Cell
Power supply
Two AA size alkaline batteries
Connectivity
Bluetooth 4.0 LE(1Mbps / 2Mbps)
Memory capacity
100 test results with test number, date and time
Battery life
Approx. 1000 tests
Dimensions(WxHxD)
60 x 128 x 24 mm
Weight
130 g including batteries
Calibration
Every 12 months or after 500 tests
Operating temperature
-5 ~ 40