User's Manual

Table Of Contents
E-45
E
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
80 MHz to 2.5 GHz
d = [3.5 / E1] P
80 MHz to 800 MHz
d = [7 / E1] P
800 MHz to 2.5 GHz
Where
P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
the recommended separation
distance in meters (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey
a
,
should be less than the compliance
level in each frequency range
b
.
Interference may occur in the
vicinity of equipment marked with
the following symbol.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects, and/or people.
Table 4: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The ConMed Linvatec VP47XX and VP47XXF Monitors are intended for use in the
electromagnetic environment specified below. The customer or the user of this VP47XX /
VP47XXF Monitor should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic environment
guidance