User's Manual

Table Of Contents
E-44
E
Portable and mobile RF
communications equipment should
be no closer to any part of the
VP47XX and VP47XXF Monitors,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended Separation
Distance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to 80 MHz
3 Vrms
150 kHz to 80 MHz
d = [3.5 / V1] P
Table 4: Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The ConMed Linvatec VP47XX and VP47XXF Monitors are intended for use in the
electromagnetic environment specified below. The customer or the user of this VP47XX /
VP47XXF Monitor should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Compliance Level
Electromagnetic environment
guidance