2.4 Recommended Annual Monitor Maintenance Requirements
Table 1: Recommended Annual Monitor Maintenance Requirements
Test Type Test Value
Ground Impedance ZG < 100 mOhm from the ground pin on the power inlet module
to the Console’s exposed metal parts. *
Test Chassis Leakage Currents IL < 100 uA in NORMAL Condition.
IL < 500 uA in Single Fault Conditions [300 uA US deviation] *
Test Earth Leakage Currents IL < 500 uA NORMAL Condition [300 uA US deviation]
IL < 1 mA Single Fault Condition *
Test Dielectric Withstand Test Line and Neutral to Ground @ V = 1500 V~, no breakdown *
* See IEC 60601-1 for test methods.