User`s manual

CM-030PMCL-RH
19
LVAL_async timing
Fig.15. Pre-select trigger mode. LVAL asynchronous
LVAL ASYNC
3μ s± 1μ s
TRIG
2L(min)
EEN
FVAL
FULL: 6L 7.1L(97.44us~115.30us) 
Vbinnin g:4L 5.2L (80.28us~104.36us)
In case the trigger pulse is input during FVAL being LOW,
LVAL async accumulation is initiated.
Exposure
CCD Exposure