User`s manual
62 Chapter 6: Sample Experiments
Calorimetry Sciences Corp.
CSC 5300 N-ITC III 63
User’s Manual
and click the burette up icon to initialize
the plunger position.) Insert the syringe and
burette drive into the N-ITC III as shown in
Figure 6-3.
Turn on the stirrer at 150 rpm, allow the sys-
tem to re-equilibrate until the heat reading on
the calorimeter is stable, and then begin the
experiment. Enter a le name for the data at
the prompt.
Results
The results should be similar
to those shown in Figure 6-4.
Note that each peak has the
same area except for the rst.
Typically the rst injection
shows less heat than expected.
This is often due to diffusion
across the tip of the needle or
to difculties in positioning
the burette drive. For 5 μL
injections of 1.00 mM HCl at
25°C, the expected heat is 237
μJ. The protonation enthalpy
in J/mol at any temperature
between 5 and 50°C is given
as:
2
nprotonatio 0.59275t -102.28t 49659- H +=∆
Equation 6-1
Figure 6-3: Inserting the burette drive into the
N-ITC III.
Figure 6-4: Typical calibration graph – Protonation of Tris
base. The last injection peak in the gure is smaller because
the titrant in the syringe was consumed.