50 Chapter 5: Software Description
Calorimetry Sciences Corp.
CSC 5300 N-ITC III 51
User’s Manual
The initial Area display is shown in Figure 5-23. If you did not enter values for Ligand,
Macromolecule, and cell volume when you ran your experiment, enter the values in the
appropriate box.
If, for some reason, a parameter of a particular injection (or injections) changed, you
can revise the value in the data spreadsheet. To change a value, highlight it, then enter
the new value in the data entry box (Figure 5-24). To recalculate the spreadsheet values,
single-click the Recalc button.
When you are satised with any additions or changes, select FileSave to write the
changes to disk.
Modeling and Fit of the Data
Once you have checked your raw experiment data for errors and entered the key experi-
ment parameters, you can model and t the data (Figure 5-25).
To t and model the data:
1. Select the desired model.
(See Modeling, for addi-
tional information on mod-
eling.)
2. Single-click the Fit Data
icon on the toolbar.
3. The computer will
calculate the values n, K,
and ∆H.
After tting the data you
can test the quality of your
t using BindWorks built-in
statistical functions.
Figure 5-25: A completed model and statistical histogram.