User`s manual

40 Chapter 5: Software Description
Calorimetry Sciences Corp.
CSC 5300 N-ITC III 41
User’s Manual
Setup Button
Used to enter the basic parameters of an experiment (Figures 5-10 and 5-11). The
parameters are as follows:
Injection or Pulse Interval
The time interval, in seconds, between injections or electrical pulses.
Injection Volume or Pulse Size
The size in µL for an injection or µJ for a pulse.
Number of Injections or Pulses
The number of injections or electrical pulses required for the experiment.
Monitor Tab
Figure 5-12 shows the Monitor tab. This tab allows the user to view data being collected
by the N-ITC III on a virtual strip chart. When you rst start the ITCRun program,
data from the ITC scrolls slowly across the screen but is not saved. Only when the
ExperimentStart button is clicked or the Go icon is chosen, will the displayed data be
saved to a le. The Monitor strip chart is used so that the user can watch this unsaved
data in order to judge when the N-ITC III is equilibrated or ready to begin an experiment.
Figure 5-10: Injection Setup.
Figure 5-11: Heater Pulse Setup.