Use Instructions
55
PROGRAMMER SYSTEM EMC PRECAUTIONS
The Model 9010 Programmer System needs special precautions regarding Electromagnetic
Compatibility (EMC) and needs to be installed and put into service according to the EMC
information provided in this guide.
Portable and mobile RF communications equipment can affect the Model 9010 Programmer
System.
The use of power cords or USB cables other than those supplied with the Model 9010
Programmer System may result in increased emissions or decreased immunity.
The Model 9010 Programmer System should not be used adjacent to or stacked with other
equipment. If such use is required, then the Model 9010 Programmer System should be
observed to verify normal operation in this configuration.
PROGRAMMER SYSTEM RF SPECIFICATIONS
The Model 9010 Programmer System may be interfered with by other equipment, even if
that other equipment complies with CISPR emission requirements. The RF telemetry
operating specifications are:
MICS band 402-405 MHz. The effective radiated power is below the limits specified in:
• USA: 47 CFR 95 Subpart I
• Canada: RSS-243
2.4 GHz band 2.4-2.4835 GHz. The effective radiated power is below the limits specified
in:
• USA: 47 CFR 15.249
• Canada: RSS-210