Instructions / Assembly

Technical Support: 1-800-721-7360, Monday-Friday, 8:00 a.m. – 4:30 p.m., CST
4. Mounting the Filter System
A. Select a location under the sink to mount the filter assembly. Choose a location
where the filter will be easily accessible.
NOTE: Allow 1
1
2-inches (38 mm) clearance below housing or 12-inch (305 mm) below
filter head to enable filter cartridge changes.
NOTE: Filter head should be mounted in vertical position, use mounting bracket as a
template to mark screw locations. Mount filter head in marked location using screws.
Filter head should be mounted on a stud or firm surface. The mounting
bracket will support the weight of the filter and help prevent strain on the cold water
line.
B. If filter assembly is mounted on a side wall next to drawers, remove drawers.
Cabinet wall may be too thin to support filter assembly. We recommend that a
3" x 3" x 1/2" wood block be used on the back side of the cabinet to allow the
screws to penetrate through the cabinet and into the wood block. This will allow
the filter assembly to be fully supported. Using the filter head assembly as a tem-
plate, drill two 1/4-inch holes in the side wall. Position the block of wood behind
these two holes and drill two 1/8-inch holes into the block. Use the two screws
provided to mount filter assembly head to side wall and on the block of wood.
5. Connecting the Supply Adapter and Inlet of Filter
A. Determine the length of plastic tubing needed to connect the inlet (left) side of the
filter with the supply adapter. Be sure to allow enough tubing to prevent kinking
and cut the tubing squarely. Place a mark 5/8-inch (16 mm) from both ends of
the tubing.
B. Wet tubing with water and insert into the supply adapter, 5/8-inch (16 mm) until
mark is flush with fitting.
C. Wet tubing with water and insert into the filter inlet, 5/8-inch (16 mm) until mark
is flush with fitting.
NOTE: Disconnecting the Tubing from the Quick-Connect Fittings. Routine maintenance
and cartridge replacement will not require that you disconnect the tubing from the filter
system; however, tubing may be quickly and easily removed from the fitting if neces-
sary. First, turn off the water supply to the filter. Open faucet, then press in the grey
collar around the fitting while pulling the tubing with your other hand.
6. Connecting the Faucet
Do not over-tighten compression nut. Use caution not to bend or crimp
tubes when securing.
A. Determine the length of plastic tubing needed to connect the outlet (right) side of
the filter with the faucet. Cut the tubing squarely.
Do not bend or crimp tube when inserting.
B. Slide plastic compression nut
A
over tubing followed by the ferrule
B
, make
sure the ferrule is oriented properly, then place insert
C
into the end of tubing.
C. Push the tubing firmly into the end of the faucet stem
D
and hand-tighten the
compression nut onto threads until secure. Then tighten approximately 1/2 turn
with a wrench.
D. Remove spare battery from side of filter cartridge and put inside the slot
E
just
below the Culligan logo. Once the battery is inserted correctly, the LED will blink
red for about 30 seconds then it will shut off. To indicate the filter needs to be
changed within the next 30 days, it will flash yellow. When the filter needs to be
changed immediately, it will flash red. The battery will need to be changed every
time a filter is changed.
7. Installing the Cartridge
Hold cartridge from the bottom when installing or changing the cartridge. Use caution
not to scrape knuckles on bracket when locking the cartridge into place. Line up the
arrow on the cartridge with unlocked padlock symbol on head. Insert cartridge and turn
arrow to locked padlock symbol. See the diagrams in Filter Cartridge Replacement on
page 4.
CAUTION:
CAUTION:
CAUTION:
3
4A
5/8"
(16mm)
5
5/8"
(16mm)
A
B
Insert
Inlet
Outlet
5C
Cabinet Wall
Wood
Block
4B
6
A
B
D
C
ferrule
orientation
E