Specifications

20
3. Slide the flow switch located on the side of the mechanical flow meter slightly upwards away
from the flow indicator inside the sight tube (mark original position before doing so). The “NO
FLOW ALARM” indicator should start blinking for 10 seconds and then illuminate constantly as
the pump is turned off. Slide flow switch back into original position and reset alarm by
pushing the “RESET ALARM” button.
Note: If any of the above described alarm test procedures fail or if any alarm trip value
deviates immediately contact AXI International.
Note: All filter elements should be replaced at least every six months.