21
Dew point
Accessories for dew point
measurement and calibration
Standard measuring
chamber for compressed
air up to 16 bar
Order no. 0699 3390
High-pressure measuring
chamber for compressed
air up to 350 bar*
Order no. 0699 3590
Measuring chamber for
granulate dryers
up to 250 mbar
Order no. 0699 3490
Measuring chamber
for atmospheric
dew point
Order no. 0699 3690
Measuring chamber
for respiratory air
bottles up to 350 bar*
Order no. 0699 3790
* in case of pressures
higher than 50 bar please
order special version
FA 400 / FA 410.
Mounting recommendation
The dew point meters can be
mounted directly into the air
stream.
However, we recommend
always to use a screwable
measuring chamber.
Screwable measuring
chamber
Advantage: Easy installation
via fast coupling
The right measuring chamber for each measuring task:
CS Service Software
for FA/VA 400 sensors including PC
connection set, USB adapter and
interface adapter to the sensor
as well as CSM-S for data recording.
The humidity sensors FA 400, FA 410,
FA 415, FA 416 can be connected to the
PC and the following adjustments can
be carried out by means of the CS
Service Software:
Scaling of the 4...20 mA
analogue output
Selecting the units: % RH, °C
td
,
g/m
3
, mg/m
3
, ppm V/V
Reading out of: Version no.,
production date, serial no.,
date of last calibration
Adjustment of the alarm limits
Single-point calibration (adjustment) -
for this purpose a reference
measuring instrument is required
By means of the CSM-S the measured
data can be read out online at the PC.
It can be stored and evaluated in graphic
and in table form.