Dew point
Dew point set DS 52 for adsorption dryers
Dew point set DS 52
consisting of:
• Digital process meter DS 52
• Dew point sensor FA 410
• Standard measuring
chamber
The measuring range of -80...20 °Ctd makes the dew point set DS 52 with alarm the
ideal dew point monitor with analogue output 4...20 mA.
The dew point sensor FA 410 is extremely long-term stable and can be installed
and removed quickly and easily under pressure by means of the screwable measuring
chamber including quick-coupling. The dew point set is supplied ready for plug-in,
i. e. everything is completely wired. The alarm values can be freely adjusted.
Description Order no.
Dew point set DS 52 for adsorption dryers consisting of: 0600 0420
DS 52 LED digital process meter in wall housing 0500 0007
FA 410 dew point sensor (-80...20 °Ctd) 0699 0410
Standard measuring chamber up to 16 bar 0699 3390
Connection cable for VA/FA Series, 5 m with M12 plug 0553 0104
Options:
Power supply 24 VDC (instead of 230 VAC) Z500 0001
Power supply 110 VAC (instead of 230 VAC) Z500 0002
Alarm unit mounted at wall housing Z500 0003
Alarm unit for external mounting with 5 m cable Z500 0004
Additional accessory:
Precision calibration at -40°Ctd including ISO certificate 0699 3396
Special features
System ready for plug-in:
Everything completely wired
No time-consuming studying of the
instruction manual
2 alarm contacts (230 VAC, 3 A)
pre- and main alarm freely adjustable
4...20 mA analogue output
Option alarm unit:
Buzzer and continuous red light
Technical data
dew point sensor FA 410
Measuring range: -80...20 °Ctd
Accuracy: ± 1 °C at 20...-20 °Ctd
± 2 °C at -20...50 °Ctd
± 3 °C at -50...-80 °Ctd
Pressure range: -1...50 bar
(with meas. chamber 16 bar),
special version up to 350 bar
Protection class: IP 65
EMV: according to DIN EN 61326
Operating temp.: -20...70 °C
Burden analogue
output: < 500 Ohm
Screw-in thread: G 1/2" without meas. chamber
Technical data
display DS 52
Dimensions: 118 x 92 x 93 mm
Display: LED red, 7 segments,
height: 13 mm, 5 digits,
2 LED for alarm relay
Keypad: 4 keys
Input: dew point sensor FA 410
Power supply: 230 VAC, 50/60 Hz;
Option: 24 VDC or
110 VAC 50/60 Hz
Alarm outputs: 2 x relay output, changeover
contact, 250 VAC, max. 3 A
Operating -10...+60 °C (storage temp.
temperature: -20 °C...+80 °C)
Alarm thresholds: freely adjustable
Hysteresis: 2 °Ctd
Analogue output: 4...20 mA = -80...20 °Ctd
Option alarm unit (buzzer
and continuous red light):