Specifications
© 2011 Cardiovascular Systems, Inc. 6 90168-00.K
Diamondback 360 Orbital Atherectomy Device (DB 360)
The DB 360 includes the following subcomponents:
• Control Handle – Supports the air turbine.
• Triple Cable with Quick Connector – includes:
o Optical Tachometer Fiber – Connects to the Controller and provides for measurement of
Shaft/Crown rotational speed;
o Turbine Gas Intake Hose – Delivers compressed gas from the Controller to the Control Handle;
o Saline tubing – Used to deliver saline from the saline bag to the DB 360. A section of the tubing is
positioned within the roller pump located on the Controller to provide the required saline flow to the
DB 360.
• Crown Control Knob – Allows for independent extension of the Crown allowing for accurate positioning.
• Drive Shaft – The drive shaft or Shaft is a flexible, helically wound, stainless steel, multiple wire shaft that drives
Crown rotation. The Shaft has a central lumen that allows for passage of a guide wire.
• Eccentric diamond-coated Crown– Attached to the drive shaft and functions as the abrasive surface. It tracks
and rotates over the guide wire, rotating at high speed to remove or reduce the occlusive tissue into microscopic
particles. Various Crown sizes are available.
• Saline Infusion Port – Allows for infusion of saline through the Sheath around the Shaft.
• Sheath – The polyethylene Sheath covers the Shaft proximal to the Crown. The Sheath protects the shunt and
vascular tissue from the spinning Shaft and permits flow of saline to the area of the abrading surface of the
Crown.
• Turbine – An internal Control Handle component, driven by compressed gas (air or nitrogen) and rotates the
Shaft/Crown.
Figure 1. Diamondback 360 Orbital Atherectomy Device (DB 360)
Triple Cable
with Quick
Co
nn
ector
Control
Handle
Crown
Control
Knob
Crown
Sheath
covering
Shaft
Saline
Tubing
Saline
infusion
port