Instruction manual
2.4 Installation / Operation
2.4.1 Installing and Operating the 2082-14x-47
1. Connect external Amplifier Plate Assembly to 2082 via MS connector, J4.(Figure 2.1).
2. Connect two 100-240 ±10% VAC, 47 - 63 Hz power cords to AC1 and AC2 on the
back panel (Figure 2.1).
3. Be sure DS8 & DS9 (green, DC Power) are on and red Alarm indicators are off (Figure 2.2).
4. Check that DS1 and DS2 are lit. If not, then either a fuse is missing or blown in F1 or F2,
respectively (Figure 2.1).
5. Set the current windows for AMP1 and AMP2 (See Section 2.5 Menu Settings).
6.
AC Fuses - The fuses are 5 mm X 20 mm, 2 amp slow blow (Type T) and are inserted in the far slot
in the drawer below the AC inputs as shown in Figure 2.3. There is a spare fuse in the near slot.
7. If a fuse continues to open, the corresponding power supply is most likely defective.
AC Fuse - 2 amp slow blow (Type T),
5 mm X 20 mm
FUSE DRAWER
SPARE FUSE
~
INPUT
100-240± 10%VA
C
47-63 Hz
2A MAX
FUSE
TYPE T 2A GDC
250 VOLT
FOR 100 - 240 V~
~
FIGURE 2.3 Fuse Location and Spare Fuse
2082-14x-47 Manual, Rev. B Page 19 12/19/13