Technical data

Preventive Maintenance 3
Agilent N5508A Hardware Reference 43
Instrument Installation
To install or re-install an instrument in a system, use one of the following
procedures.
Required tools
#2 Phillips screwdriver
#2 POZIDRIV screwdriver
Half-Rack-Width instrument
To install the instrument in a rack
Step Note
1 Make sure the system is powered off. For details, see the system installation guide
or system users guide.
2 Re-attach the lock link that secures the front
of the returned instrument to its partner
half-rack-width instrument.
Use a #2 POZIDRIV screwdriver.
See Figure 11 on page 42.
3 Re-attach the lock link that secures the rear
of the instruments together.
Use a #2 POZIDRIV screwdriver.
4 Insert the attached instruments in the same
slot from which you removed them, sliding
them along the support rails until they meet
the rack-mount ears.
The rack-mount ears stop the instruments at
the correct depth.
5 Replace the rack panel in front of the
instruments and secure the four corner
screws.
The screws are located near the corners of
the face of the instrument.
Use a #2 Phillips screwdriver.
6 Confirm that the instrument is turned off.
7 Connect the appropriate cables to the
instruments (front and rear), including the
power cords.
8 Power on the system. For details, see the system installation guide
or system users guide.